For long-term storage of biological samples, cell lines or valuable research material, cryogenic storage comes into play. The two most commonly used methods are storage with liquid nitrogen (LN2) and storage in a mechanical cryogenic freezer. Which method suits your lab best depends on safety, capacity, ease of use and logistics. In this article we […]
For long-term storage of biological samples, cell lines or valuable research material, cryogenic storage comes into play. The two most commonly used methods are storage with liquid nitrogen (LN2) and storage in a mechanical cryogenic freezer. Which method suits your lab best depends on safety, capacity, ease of use and logistics. In this article we compare both options, with practical input from Robert van der Kuijlen, director of Poly Temp Scientific.
Cryogenic storage covers all storage below a temperature of -130 °C. Below this threshold, all biological processes in living cells come to a standstill. The ageing process stops and cellular material can be preserved for virtually indefinite periods without loss of biological integrity. This makes cryogenic storage the standard for cryopreservation of cells, tissue, sperm, egg cells, embryos and other biological samples in research, clinical biobanks and pharmaceutical development.
The term cryogenic comes from the Greek (kryos = icy cold, genes = producer of) and refers to extremely low temperatures, typically below -150 °C. Cryogenic gases are gases that become liquid through strong cooling and are stored or transported in that liquid form. Well-known examples include liquid nitrogen (-196 °C), liquid argon (-186 °C), liquid oxygen (-183 °C), liquid hydrogen (-253 °C) and liquid helium (-269 °C).
The liquid nitrogen temperature in the liquid phase is -196 °C (77 Kelvin). This is the boiling point of nitrogen at atmospheric pressure. As long as liquid is present, the temperature stays stable around this point through evaporative cooling. Above the liquid level, in the so-called vapour phase, the temperature lies between approximately -150 °C and -190 °C, depending on position in the tank and fill level.
By comparison: a mechanical cryogenic freezer maintains the full storage volume at a uniform -150 °C. This sits just below the glass transition temperature of water (-130 °C), keeping cells safely in cryogenic rest without ice crystal formation or metabolic activity occurring. The liquid nitrogen temperature therefore offers a wider safety margin, while the mechanical cryogenic freezer works closer to the cryogenic threshold without ever exceeding it.
Liquid nitrogen has been used for cryogenic storage for decades and remains the standard solution in many labs. The nitrogen is stored in a vacuum-insulated dewar or nitrogen tank, such as the CryoExtra cryogenic storage vessel, which maintains the extremely low temperature through minimal heat exchange with the environment.
A dewar, also called a dewar vessel, is a vacuum-insulated vessel specifically designed for storage and transport of liquid nitrogen and other cryogenic liquids. The double-walled construction with vacuum between the walls minimises heat transfer. In addition, modern dewars often contain an aluminium inner wall and advanced insulation layers that further limit static evaporation. A good dewar can hold liquid nitrogen for weeks without refilling, depending on capacity and neck opening.
For cryogenic storage with LN2 there is a distinction between automatically and manually filled tanks:
Working with liquid nitrogen carries serious safety risks. Asphyxiation is the main risk: 1 litre of LN2 produces approximately 700 litres of nitrogen gas upon evaporation, which displaces oxygen. In poorly ventilated spaces this can be life-threatening within seconds. In addition, direct contact causes cold burns (cryogenic burns), and dangerous pressure build-up can occur in closed lines.
Another practical disadvantage: liquid nitrogen evaporates continuously, even in well-insulated dewars. That means recurring deliveries, logistical dependence on the gas supplier and an ongoing consumption cost. Some organisations decide to abandon LN2 entirely for these reasons.
For laboratories choosing LN2 storage, Poly Temp supplies various Thermo Scientific cryopreservation systems. The CryoPlus LN2 storage system is an auto-fill vapour phase system with microprocessor-controlled level monitoring and suits organisations choosing automatically filled storage. For controlled freezing of cells before transition to long-term cryogenic storage, the CryoMed controlled rate freezer uses dual solenoid LN2 injection for precise freezing profiles and optimal cell viability.
A mechanical cryogenic freezer is an electrically cooled freezer that reaches a uniform temperature of -150 °C through multi-stage cascade cooling without liquid nitrogen. An example in the Poly Temp range is the Labselect PTSL200CF -150 °C cryo freezer: a chest freezer with dual compressor and four-stage cascade system, running on HC refrigerants (GWP near 0) and standard 230V (no three-phase power needed). With a capacity of approximately 200 litres, built-in battery back-up and extensive alarm functions, the PTSL200CF offers reliable cryogenic storage for laboratories, biobanks and R&D environments that want to avoid the risks and logistics of LN2.
Full utilisation of storage volume. Robert van der Kuijlen (director Poly Temp) explains:
With liquid nitrogen the user can store in the vapour phase, but that costs storage capacity. With mechanical cooling that doesn’t apply. The 200 litres available can be fully used for samples. Getting material in and out of our -150 °C freezer is also easier than from an LN2 vessel. The freezer is only one metre tall.
The table below summarises the main differences.
| Aspect | Liquid nitrogen (LN2) | Mechanical cryogenic freezer |
|---|---|---|
| Temperature | -196 °C (liquid) / -150 to -190 °C (vapour) | -150 °C uniform |
| Safety margin during power outage | Passive, no electricity required | 20 °C margin to -130 °C |
| Asphyxiation risk | Yes, ventilation and O2 detection required | No |
| Cold burns | Risk during filling and handling | Minimal |
| Consumption | Continuous evaporation, periodic delivery | Electricity only |
| Storage capacity utilisation | Limited by vapour or liquid phase choice | Fully usable |
| Height / ergonomics | Tall dewar, harder to load | Approximately 1 metre, user-friendly |
| Logistical dependence | Nitrogen deliveries, gas supplier contract | Only mains power and service |
| Maintenance | Filling, level monitoring, dewar maintenance | Periodic service interval |
The choice between LN2 and mechanical cryogenic cooling depends on several factors. You can also view our full range of cold storage equipment for laboratories for a broader overview of options.
Unsure which cryogenic storage suits your situation? Our specialists are happy to help you think through capacity, safety and infrastructure. Contact us for personal advice or request a quote for your preferred model.