40+ jaar expertise in laboratoriumapparatuur
Eigen servicedienst en onderhoud
ISO 9001 gecertificeerd

n-Laf NPU cage changing station

Delivery timeiFor the current delivery time, please contact us by phone at +31 515 575 105 or send us an email at info@polytemp.com.

Cage changing station with HEPA H14 laminar flow for safe and efficient changing of rodent cages. Animal, operator and environmental protection, electrically adjustable and fully custom built.

Laminar flow from HEPA H14 filters for a clean, animal-safe environment
Ventilated work table and rear wall protect the operator against dust
Electrically height adjustable work table for optimal ergonomics
Automatic cage dispensers for clean and dirty cages on both sides
Custom design available, incl. emptying station and stainless steel AISI304/316
40+ years of expertise In-house service team Fast delivery in the Netherlands

Description

The n-Laf NPU cage changing station from Nordic Labtech is a cage changing station for efficient and safe changing of rodent cages. Thanks to the laminar airflow from H14 filters, a clean environment is created above the work surface that protects the animals against contamination. At the same time, the ventilated work table protects the operator against dust and allergens from the cages. In this way, this animal cage changing station combines animal, operator and environmental protection in one compact system.

Ergonomics and daily workflow

The work table is electrically height adjustable, so every operator can work in an optimal, ergonomic posture. On either side of the table there are automatic cage dispensers: a stack of clean cages on one side and a stack of dirty cages on the other. This allows cage changing to run in a logical, fast movement without unnecessary lifting or turning. Behind the dispensers the rear wall is ventilated, which gives the operator extra protection against dust that is released when opening and closing the cages.

Protection for animal, operator and environment

The laminar flow from the HEPA H14 filters keeps the zone above the cages clean, while the ventilated work table and rear wall extract the dust towards the operator protection. This design makes the cage changing station suitable for facilities where both the health of the animals and the safety of the staff are paramount. The latest fan and filter technology also ensures a low noise level and low energy consumption during continuous use.

Custom design and configurations

The station has a standard working area of 2000 x 1400 mm, but is supplied as a custom solution tailored to your room and workflow. On request we adapt the system into an emptying station within the LAF zone, protecting the operator and the environment against dust when emptying old bedding. For extra hygiene and cleanability, a version in stainless steel AISI304 or AISI316 is optionally available. Request a no-obligation quote for a custom laminar flow solution for your animal facility.

Specifications

Specification n-Laf NPU cage changing station
Product type Cage changing station with laminar airflow
Brand Nordic Labtech
Model NPU Cage Changing Station 2000×1400
Protection Animal, operator and environment
Material White powder coated steel, optionally stainless steel AISI 304 or AISI 316
Worktable With or without, fixed height or electrically elevated, optionally ventilated with process air
Cage dispensers Optional
Exterior dimensions (W x D x H) 2000 x 1900 x 2700 mm
Interior dimensions (W x D x H) 2000 x 1400 x 2000 mm
Downflow velocity Optimum 0.30 m/s, variable from 0.1 to 0.6 m/s
Average downflow 0.1 to 0.6 m/s plus/minus 10 %
Downflow volume Typically 3000 m3/h
Total exhaust air Typically 500 m3/h
Main filter HEPA H14 to EN 1822
Pre-filter G3 or G4
Laminator PETP, pressure drop max. 40 Pa at 0.5 m/s
Noise level Less than 56 dB(A) at 0.30 m/s, less than 50 dB(A) at reduced speed
Fans EBM or corresponding
Lighting Very low energy consumption LED, min. 800 lux or more if required
Power supply 230 VAC, protected with 10 A or 16 A
Energy consumption with clean filters 0.4 kW
Type n-Laf NPU cage changing station
Brand Nordic Labtech
Application Changing rodent cages in animal facilities
Standard size 2000 x 1400 mm
Filter type HEPA H14
Down flow speed 0.30 m/s (adjustable 0.1 – 0.6 m/s)
Downflow volume approx. 3000 m3/h
Exhaust volume approx. 500 m3/h
Work table Electrically height adjustable
Cage dispensers Automatic, on both sides (clean and dirty)
Rear wall Ventilated for operator protection
Protection Animal, operator and environment
Material Stainless steel AISI304 or AISI316 optional
Custom design Yes, incl. optional emptying station

Frequently asked questions

What is a cage changing station used for?

A cage changing station is used to change rodent cages efficiently and safely. The laminar airflow from the H14 filters keeps the zone above the cages clean and protects the animals, while the ventilated work table and rear wall protect the operator against dust and allergens.

Does this animal cage changing station protect both the animals and the operator?

Yes. The HEPA H14 laminar flow provides a clean environment above the cages for the animals, while the ventilated work table and ventilated rear wall extract the dust and thus protect the operator and the environment. The system offers animal, operator and environmental protection in one.

Can the station be adapted into an emptying station?

Yes. The cage changing station can be custom built as an emptying station in the LAF zone. This protects the operator and the environment against dust released when emptying old bedding.

Which sizes and configurations are possible?

The standard size is 2000 x 1400 mm, but the station is custom built and tailored to your room and workflow. A version in stainless steel AISI304 or AISI316 is optionally available for extra hygiene and cleanability.

Is the station energy efficient and quiet in use?

Yes. Thanks to the latest fan and filter technology, the station operates with a low noise level and low energy consumption, even during continuous use in an animal facility.

00