Mobile cage changing unit for the safe changing of rodent cages. EN 12469 compliant, H14 filtration, dual ventilated work surfaces and optimal ergonomics. Available in 1045 and 1345 mm working width. Price on quote.

EN 12469 compliant with inflow greater than 0.42 m/s with both sides open
Dual ventilated work surfaces prevent left-right cross contamination
H14 filtration on supply and exhaust, optional carbon filter against gases and odours
Very quiet (less than 52 dB(A)) and light (less than 200 kg), mobile on low noise castors
Electrically adjustable base per user and dimmable LED 0-1500 Lux
40+ jaar expertise Eigen servicedienst Snelle levering NL

Description

n-Safe MCCU: safe cage changing with optimal ergonomics

The n-Safe MCCU is a mobile cage changing unit for the safe changing of rodent cages. This cage changing station combines full operator protection with well thought out ergonomics, so animal care staff work comfortably and without exposure to allergens or pathogens. As a mobile cage changing unit, it brings protection to the animals instead of moving the animals to a fixed workstation.

Full operator protection and EN 12469 compliant

The MCCU provides full operator protection and is EN 12469 compliant. Even with both sides open, the inflow remains higher than 0.42 m/s, so allergens and aerosols cannot escape to the room. The down flow of 0.28 m/s (adjustable between 0.25 and 0.38 m/s) creates a stable air curtain above the work surface. Dual ventilated work surfaces prevent cross contamination between the left and right side, which is essential when keeping animal lines and studies separate.

H14 filtration and optional activated carbon filter

Both supply and exhaust air are filtered with H14 filters, for the highest assurance against particles and biological agents. An optional activated carbon filter on the exhaust is available to capture carcinogenic gases and odours. The prefilter cassettes are easy to exchange, which keeps maintenance fast and safe.

Ergonomics, mobility and comfort in use

Weighing less than 200 kg, this cage changing station is light and mobile on low noise castors, one pair of which is lockable. The base is electrically height adjustable and programmable per user, so every operator works at the correct height. The dimmable LED lighting is adjustable from 0 to 1500 Lux. This animal facility workstation is also very quiet at less than 52 dB(A), which reduces stress for both animal and operator.

Two working widths and practical options

The n-Safe MCCU is available as the MCCU 1300 with 1045 mm working width and as the MCCU 1600 with 1345 mm working width. Optional accessories include an autoclavable stainless steel feed dispenser and a battery backup, which keeps the unit running without mains power during transport or a power outage. Worldwide remote support ensures fast assistance from a distance.

Specifications

Specification MCCU 1300 MCCU 1600
Working width 1045 mm 1345 mm
External dimensions (w x d x h) approx. 1295 x 795 x 1935-2285 mm approx. 1595 x 795 x 1935-2285 mm
Down flow rate 710 m3/h 920 m3/h
Exhaust rate 2060 m3/h 2350 m3/h
Weight 195 kg 220 kg
Down flow velocity 0.28 m/s (0.25-0.38) 0.28 m/s (0.25-0.38)
Inflow (both sides open) > 0.42 m/s > 0.42 m/s
Supply and exhaust filtration H14 H14
Optional exhaust filter Activated carbon filter Activated carbon filter
Noise level < 52 dB(A) < 52 dB(A)
LED lighting Dimmable 0-1500 Lux Dimmable 0-1500 Lux
Base Electrically adjustable, programmable per user Electrically adjustable, programmable per user
Mobility Low noise castors, 1 pair lockable Low noise castors, 1 pair lockable
Standard EN 12469 compliant EN 12469 compliant
Options Stainless steel feed dispenser (autoclavable), battery backup Stainless steel feed dispenser (autoclavable), battery backup

Frequently asked questions

What do I use an n-Safe MCCU cage changing unit for?

The n-Safe MCCU is designed for the safe changing of rodent cages. The mobile cage changing station protects the operator against allergens, aerosols and pathogens while animals are transferred to clean cages.

How does the MCCU prevent cross contamination between studies?

The work surfaces are dual ventilated, so the left and right side remain aerodynamically separated. This prevents cross contamination between animal lines or studies handled at the same time.

Does the unit meet safety standards?

Yes, the n-Safe MCCU is EN 12469 compliant. Even with both sides open, the inflow stays higher than 0.42 m/s, and supply and exhaust air are filtered with H14 filters for full operator protection.

Can the unit capture carcinogenic gases and odours?

For applications with carcinogenic gases or unpleasant odours, an optional activated carbon filter on the exhaust is available. This filter captures volatile substances and odours on top of the standard H14 filtration.

Which working widths are available and what does the MCCU cost?

The MCCU is available as the 1300 with 1045 mm working width and as the 1600 with 1345 mm working width. Prices depend on configuration and options. Request a no obligation quote for a tailored price.