Thermo Scientific TFT-80.4 fixed-angle rotor, 10 x 4.4 mL, for the Sorvall WX 80+ / WX 90+ / WX 100+ range. Rotor maximum: 80,000 rpm and 468,246 × g. Actual limits depend on the complete instrument and labware configuration.
The Thermo Scientific TFT-80.4 is a fixed-angle rotor with a nominal capacity of 10 x 4.4 mL. The rotor catalogue number is 54356. Its rotor angle is 25°. The rotor’s nominal maximum speed is 80,000 rpm. The centrifuge and tube combination may impose a lower limit; nominal capacity is not the permitted filling volume for every tube.
A fixed-angle rotor holds tubes at a fixed angle during centrifugation. This geometry is used for pelleting and, depending on the method, density-gradient separations. Working fill volume and separation conditions depend on the selected tube and protocol.
Sorvall WX 80+, Sorvall WX 90+, Sorvall WX 100+.
The rotor is listed for this instrument family. This does not mean every instrument can achieve the rotor’s maximum speed. A lower-speed centrifuge remains limited to its own rated speed; approved rotor settings, sample density and labware may impose further restrictions.
Use the manufacturer rotor guide in Downloads to identify tube sizes, working fill volumes, closures and any required adapters. Include the instrument model, rotor catalogue number and sample requirements with your request. Poly Temp confirms the configuration, supplied package and current availability before ordering.
Compare Sorvall WX+ rotors by rotor geometry, tube requirements and centrifuge model.
| Catalogue number | 54356 |
|---|---|
| Model | TFT-80.4 |
| Type | fixed-angle rotor |
| Nominal capacity | 10 x 4.4 mL |
| Instrument family | Sorvall WX 80+ / WX 90+ / WX 100+ |
| Maximum rotor speed | 80,000 rpm |
| Maximum rotor RCF | 468,246 × g |
| K factor (rotor guide) | 25.8 |
| Product weight | 3.5 kg |
No. This rotor is listed for Sorvall WX 80+ / WX 90+ / WX 100+. The exact instrument, mounting system and labware must be checked.
Not always. Nominal capacity and permitted fill volume can differ with tube material, closure and application. Follow the tube-specific manufacturer limits.
The quotation specifies the supplied parts. Tubes, closures, adapters and installation components may need to be ordered separately.