The Thermo Scientific Thermolyne premium large muffle furnace combines a spacious 14 liter working chamber with advanced temperature control up to 1200 °C. As a result, this model suits laboratories that need both larger sample volumes and high temperature precision. Specifically, stability is ± 0.3 °C and uniformity is ± 2.2 °C at 1000 °C, considerably more accurate than smaller benchtop variants. Furthermore, built-in safety systems and a Platinel II thermocouple ensure reliable performance during extended operation at high temperatures.
The most distinctive feature of this large muffle furnace is the four-heating-element configuration. Specifically, elements are placed on the top, bottom, and sides of the chamber, rather than only on the sidewalls as in simpler models. As a result, heat distributes evenly across the entire sample volume, with uniformity of ± 2.2 °C at 1000 °C. Notably, this configuration prevents hotspots and cold zones, which is critical for gravimetric analysis and reproducible thermal treatments.
Specifically, you choose from three PID microprocessor controllers depending on the complexity of your application:
Furthermore, all three models feature a Platinel II thermocouple, suited for extended operation at temperatures up to 1400 °C.
Specifically, this large muffle furnace is found in laboratories working with larger sample volumes or processing multiple samples at once. Common applications include:
In addition, this model is used in pharmaceutical development, materials research, and the food industry for combustion analysis of additives and residues.
Moreover, this large muffle furnace features multiple safety layers. Specifically, the door switch interrupts power to the heating elements when the door opens. Furthermore, thermocouple break protection cuts heating when a sensor fails, which prevents uncontrolled temperature rise. Notably, the mechanical OTP relay provides an independent layer of overheating protection. In addition, the built-in vent port removes fumes and moisture directly, extending the life of elements and the unit. Specifically, a 0.8 cm diameter port at the rear allows independent temperature measurement.
Each premium large muffle furnace ships with a Platinel II thermocouple, mechanical OTP relay, and hearth tray. As a result, the unit is immediately ready for applications up to 1200 °C without additional investment. Furthermore, a stainless steel shelf is optionally available (SH408X1, rated up to 900 °C), which doubles load capacity by allowing two sample levels. In addition, a set of hearth trays (PHX1, up to 9 pieces in a 3 × 3 pattern) and a stainless steel exhaust kit (AY408X1A, 2.5 inch ID, 5 ft length) are available.
Note: all EU models require hardwiring by a qualified technician. Specifically, no plug or power cord is supplied, because the 4400 W power demands a fixed electrical connection.
| Feature | Specification |
|---|---|
| Type | Premium large muffle furnace |
| Capacity | 14 liter (0.5 cu. ft.) |
| Temperature range | 100 to 1200 °C |
| Stability at 1000 °C | ± 0.3 °C |
| Uniformity at 1000 °C | ± 2.2 °C |
| Heating elements | Four elements, top, bottom, and sides |
| Thermocouple | Platinel II (rated up to 1400 °C) |
| Safety | Door switch, thermocouple break protection, mechanical OTP relay |
| Vent port | Built-in for fume and moisture removal |
| Monitoring port | 0.8 cm diameter, rear of chamber |
| Voltage | 240V, 50/60 Hz |
| Power | 4400 W |
| Current | 18.3 A |
| Connection | No plug, requires hardwiring by a technician |
| Certification | CE |
| Interior dimensions (W × D × H) | 33 × 25 × 18 cm |
| Exterior dimensions (W × D × H) | 48.5 × 51 × 53.3 cm |
| Shipping weight | 60.8 kg |
| Included as standard | Platinel II thermocouple, mechanical OTP relay, hearth tray |
| Optional accessories | Stainless steel shelf SH408X1 (up to 900 °C), hearth tray set PHX1, stainless steel exhaust kit AY408X1A |
| SKU | Controller | Programs | OTP relay | RS-232 |
|---|---|---|---|---|
| F6020C-33 | Single setpoint with ramp (B1) | None | Yes | No |
| F6020C-33-80 | Programmable (C1) | 1 program, 8 segments | Yes | No |
| F6030CM-33 | Programmable (D1) | 5 programs, 16 segments | Yes | Yes |
The Thermo Scientific Thermolyne premium large muffle furnace provides a 14 liter working chamber for laboratories handling larger sample volumes up to 1200 °C. Specifically, four heating elements positioned on the top, bottom, and sides deliver excellent temperature uniformity of ± 2.2 °C at 1000 °C. Furthermore, three PID controllers are available, from single setpoint with ramp to programmable with RS-232 communications. In addition, an optional stainless steel shelf is available that doubles load capacity for applications up to 900 °C.








A large muffle furnace is a laboratory furnace with a substantial chamber capacity, in this case 14 liter, that heats samples to high temperatures inside an insulated combustion chamber. Specifically, a larger size enables treating multiple samples at once or processing larger parts. In addition, premium models such as this one deliver better temperature uniformity through multiple heating elements around the chamber.
The Thermolyne benchtop muffle furnace offers 5.8 liter, while this premium large muffle furnace delivers 14 liter: nearly 2.5 times as much. Moreover, this model features four heating elements around the chamber instead of two on the sides, which results in better uniformity (± 2.2 °C versus ± 3.6 °C at 1000 °C). In addition, this model includes a Platinel II thermocouple suited for temperatures up to 1400 °C as standard, while the smaller benchtop models use a Type K thermocouple.
Specifically, the choice depends on your process. For applications with a controlled heatup curve and a hold period, the F6020C-33 (B1) suffices. By contrast, the F6020C-33-80 (C1) offers a programmable profile of 8 segments, ideal for applications with multiple temperature steps. Furthermore, the F6030CM-33 (D1) is the choice when you want to store 5 different programs or log data via RS-232 communications.
The premium large muffle furnace operates at 4400 W, which is significantly higher than smaller models. As a result, this furnace requires a fixed electrical connection through hardwiring by a qualified technician. Specifically, no plug or power cord is supplied, because standard plugs cannot safely deliver this power level during continuous use.
Yes, with the optional stainless steel shelf (SH408X1) you create a second loading level within the same chamber. As a result, load capacity effectively doubles. In addition, it is important to know: this stainless steel shelf is rated for applications up to a maximum of 900 °C, so for higher temperatures you must fall back to a single loading level on the standard hearth tray.