Thermo Scientific Wellwash microplate washer

Article no.: 5165000 / 5165040Delivery timeiFor the current delivery time, please contact us by phone at +31 515 575 105 or send us an email at info@polytemp.com.

The Thermo Scientific Wellwash is a straightforward microplate washer for reproducible routine ELISA protocols using 96-well plates and strips. Available with a 1 x 8 or 1 x 12 wash head.

Select the supplied wash-head configuration. The selected catalogue number will be included in your quotation request.

For 96-well ELISA plates and strips
Choice of 1 x 8 or 1 x 12 wash head
Residual volume below 2 µL
Up to 99 stored protocols
Safe non-pressurised bottles
40+ years of expertise In-house service team Fast delivery in the Netherlands

Description

Reliable 96-well microplate washing for routine ELISA

The Thermo Scientific Wellwash microplate washer automates aspiration, dispensing, washing and linear shaking of 96-well microplates and strips. It is designed for laboratories running a limited number of similar ELISA protocols routinely and seeking to reduce manual variation.

The large colour display and graphical user interface provide control over wash cycles, soak time, shaking speed, dispensing height, aspiration height and aspiration speed. Up to 99 protocols can be stored. USB supports protocol transfer between instruments and export of reports and log files.

Low residual volume and adjustable aspiration

Crosswise sweep aspiration operates at two positions across the well bottom, resulting in a specified residual volume below 2 µL. Adjustable wash parameters allow the process to be matched carefully to the plate and assay.

Safe and straightforward liquid handling

The 2 L wash bottle and 2 L waste bottle are not pressurised. Liquid-level sensors monitor both bottles, a plate sensor checks that a microplate is present and the supplied aerosol cover helps limit aerosol spread. Automatic prime and rinse functions support consistent startup and a clean liquid path.

When should you choose standard Wellwash?

  • When washing 96-well plates or strips only.
  • When one wash buffer is sufficient.
  • For recurring, similar ELISA protocols.
  • When cell washing, 384-well plates and robotic integration are not required.

If you also process 384-well plates, require multiple buffers, wash adherent cells or use laboratory automation, see the Thermo Scientific Wellwash Versa microplate washer.

Use: for research use only; not for use in diagnostic procedures.

Accessories and replacement parts

View all 24 original wash heads, bottles, stands, aerosol covers and service parts for Wellwash and Wellwash Versa.

View all Wellwash accessories

Specifications

Plate formats 96-well plates and strips
Buffer positions 1
Wash head 1 x 8 or 1 x 12, depending on model
Wash volume 50-1,000 µL
Dispense volume 50-400 µL
Prime/rinse volume 5-100 mL
Residual volume <2 µL
Dispensing accuracy Less than 5% error at 300 µL
Dispensing precision <3% CV at 300 µL
Protocols Up to 99
Connection USB; no robotic integration
Dimensions (W x D x H) 38.5 x 24 x 34.5 cm
Weight 8 kg
Power 100-240 V, 50/60 Hz

Models

Catalogue number Model Included
5165000 Wellwash 1 x 8 1 x 8 wash head, 2 L wash bottle, 2 L waste bottle and aerosol cover
5165040 Wellwash 1 x 12 1 x 12 wash head, 2 L wash bottle, 2 L waste bottle and aerosol cover

Frequently asked questions

Which plates can standard Wellwash process?

Standard Wellwash is designed for 96-well microplates and strips. A Wellwash Versa with the optional 1 x 16 wash head is required for 384-well plates.

What is the residual volume?

The specified residual volume is below 2 µL due to crosswise sweep aspiration at two positions across the well bottom.

Which wash heads are available?

Standard Wellwash is available as a 1 x 8 model (5165000) and a 1 x 12 model (5165040).

How many protocols can Wellwash store?

Up to 99 wash protocols. Protocols, reports and log files can be transferred through USB.

Can Wellwash wash cells or connect to robotics?

No. Select Wellwash Versa when cell washing or laboratory automation integration is required.

Is Wellwash intended for diagnostic use?

No. The manufacturer specifies the instrument for research use only.