Application and configuration
The n-Safe B2 is a Class II B2 safety cabinet with 100% exhaust to an engineered building system. It is intended for processes where recirculation is unsuitable and always requires facility engineering.
Protection and versions
- Protects: product, operator and environment within the confirmed process and exhaust conditions.
- Widths: 1200 / 1500 / 1800 mm.
- Standard: 100% exhaust, no internal exhaust fan and G4 prefilter.
- Brochure passive-exhaust values: approximately 1070 / 1320 / 1590 m³/h for the respective widths.
Configurable options
| Group | Possibilities | Status |
|---|---|---|
| Stand | Fixed, manually adjustable or electric | Option |
| Worktop | Segmented/one-piece; AISI 316; feed-throughs | Option/project |
| Media | Electricity, water, vacuum and gas | Option/project |
| UV | Integrated UV version | Option |
| Exhaust | Fixed duct, anti-blowback, alarm/interlock and building connection | Mandatory project component |
| Validation | IQ/OQ and project-specific exhaust/performance verification | Project option |
Technical conditions
Fixed building exhaust, make-up air, pressure control, capacity, failure scenario, interlock and room balance must be designed and confirmed before ordering. Total exhaust does not automatically make a gas- or vapour-producing process safe; substance properties, quantity and exposure limits remain decisive.
Application limits
This cabinet is not a plug-and-play solution and requires suitable building exhaust. Do not use it as a generic replacement for a fume hood when the chemicals are unknown.
Advice and quotation
Have the process and HVAC reviewed together. Poly Temp will coordinate cabinet width, exhaust capacity, interlocks, media and validation with the building system.