n-SAFE B2 | Class II B2 Biological Safety Cabinet

Article no.: 311 4000 1200Delivery timeiFor the current delivery time, please contact us by phone at +31 515 575 105 or send us an email at [email protected].

The n-Safe B2 is a Class II B2 safety workstation with 100% exhaust without recirculation. The system requires connection to an external extraction system. The cabinet complies with NSF/ANSI 49 and EN 12469. Available in 1200, 1500 and 1800 mm working widths.

Full service incl. installation
Quote on request
Configure accessories

Select a model to see matching accessories. Quantities apply per instrument. Options without a confirmed price can be requested in a quotation.

Price on request.

With hooks available in two lengths. Price on request.

Price on request.

This product is supplied including installation. Request a quotation for a tailored price, optionally extended with additional options.
Class II B2 with 100% exhaust
No recirculating air in the workspace
Compliant with NSF/ANSI 49 and EN 12469
HEPA H14 main and exhaust filtration
Available in 1200, 1500 and 1800 mm working widths
40+ years of expertise In-house service team Fast delivery in the Netherlands

Description

n-Safe B2 Class II B2 Safety Workstation

100% Exhaust Without Recirculation

The n-Safe B2 is a Class II B2 microbiological safety workstation according to NSF/ANSI 49. The cabinet provides protection for product, operator and environment. The system operates with 100% exhaust and has no recirculating air. This makes it suitable for applications where complete exhaust is required.

The B2 version requires connection to an external extraction system. The cabinet does not have a built-in exhaust fan. Exhaust air is passively discharged via a ducting system. The design complies with EN 12469 for product and operator protection.

The vertical downflow is 0.28 m/s and is adjustable between 0.25 and 0.53 m/s. The inflow is 0.45 m/s and is adjustable between 0.45 and 0.55 m/s. The deviation remains within ±8 percent. The system is equipped with HEPA H14 main and exhaust filters according to EN 1822.

Key Features

  • Class II B2 safety workstation according to NSF/ANSI 49
  • Protection compliant with EN 12469
  • 100% exhaust without recirculation
  • Requires external ducting, no built-in exhaust fan
  • HEPA H14 filtration according to EN 1822
  • Dimmable LED lighting 0-2000 lux
  • Front window opening 200 mm, adjustable 160-250 mm
Property Specification
Type Class II B2 microbiological safety workstation
Standards NSF/ANSI 49, EN 12469
Downflow air velocity 0.28 m/s (adjustable 0.25-0.53)
Inflow air velocity 0.45 m/s (adjustable 0.45-0.55)
Deviation ±8%
Downflow capacity 680-1010 m³/h (model dependent)
Exhaust flow rate 1070-1590 m³/h (passive, ducting required)
Noise level 46-47 dB(A) according to EN 12469
Power From 100 W
Filter technology HEPA H14 EN 1822, 99.999% at 0.3 µm
Pre-filter G4
Lighting LED 0-2000 lux, dimmable
Mains voltage 220-240 V / 50-60 Hz or 110-120 V / 50-60 Hz
Front window opening 200 mm (adjustable 160-250 mm)
Working height fixed stand 800 or 900 mm (±25 mm)
Working height adjustable 700-1050 mm
Working height electric 750-1050 mm
Minimum clearance 800 mm
 
SKU Model Outer dimensions excl. stand (W x D x H) Inner dimensions (W x D x H) Weight Downflow / Exhaust (m³/h)
311 4000 1200 n-Safe B2 1200 1303 × 799 × 1405 mm 1200 × 520(625) × 700 mm 235 kg 680 / 1070
311 4000 1500 n-Safe B2 1500 1603 × 799 × 1405 mm 1500 × 520(625) × 700 mm 265 kg 840 / 1320
311 4000 1800 n-Safe B2 1800 1907 × 799 × 1405 mm 1800 × 520(625) × 700 mm 295 kg 1010 / 1590
311 4000 2100 n-Safe B2 2100 On request On request On request On request

Specifications and availability for project-specific configurations are provided on request. Dimensions and performance depend on the configuration; data for other models do not automatically apply.

Airflow and window opening are set for the selected configuration during commissioning. Use the cabinet in its validated operating setting. Protection depends on the configuration, risk assessment, installation, testing and correct use.

Manufacturer brochure

Frequently asked questions

What characterises the n-Safe B2?

It is a Class II B2 cabinet with 100% exhaust, without recirculating air in the work area.

Why is 100% exhaust important?

No air recirculates in the work area, ideal for volatile substances.

Which filtration is used?

HEPA H14 main and exhaust filtration.

Which standards does the cabinet meet?

The cabinet meets NSF/ANSI 49 and EN 12469.

Which working widths is the B2 available in?

The n-Safe B2 is available in 1200, 1500 and 1800 mm working width.

Who is the B2 suited for?

For laboratories working with volatile or odorous substances that need full exhaust.

00
Configuration and application

Select the correct version and options

The exact version is confirmed with the quotation. Safety conditions always take priority over accessories.

Application and configuration

The n-Safe B2 is a Class II B2 safety cabinet with 100% exhaust to an engineered building system. It is intended for processes where recirculation is unsuitable and always requires facility engineering.

Protection and versions

Configurable options

GroupPossibilitiesStatus
StandFixed, manually adjustable or electricOption
WorktopSegmented/one-piece; AISI 316; feed-throughsOption/project
MediaElectricity, water, vacuum and gasOption/project
UVIntegrated UV versionOption
ExhaustFixed duct, anti-blowback, alarm/interlock and building connectionMandatory project component
ValidationIQ/OQ and project-specific exhaust/performance verificationProject option

Technical conditions

Fixed building exhaust, make-up air, pressure control, capacity, failure scenario, interlock and room balance must be designed and confirmed before ordering. Total exhaust does not automatically make a gas- or vapour-producing process safe; substance properties, quantity and exposure limits remain decisive.

Application limits

This cabinet is not a plug-and-play solution and requires suitable building exhaust. Do not use it as a generic replacement for a fume hood when the chemicals are unknown.

Advice and quotation

Have the process and HVAC reviewed together. Poly Temp will coordinate cabinet width, exhaust capacity, interlocks, media and validation with the building system.